Kyungmin Seo    Hyunsoo Kim    Yeongduk Kim    Hyeyoung Lee    Jaison Lee    Moo Hyun Lee,11footnotetext: Corresponding author.    Jungho So    Sangcheol Yoon    and Youngsoo Yoon
Abstract

The radiation emitted from radon is a critical background in rare event search experiments conducted at the Yemi Underground Laboratory (Yemilab) in Jeongseon, Korea. A Radon Reduction System (RRS) has been developed and installed in Yemilab to reduce radon concentration in the air. The RRS primarily provides a purified air of 50m3/h50superscriptm3h\rm 50\ m^{3}/h50 roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT / roman_h to the cleanroom used to assemble crystal detectors in the AMoRE, a neutrinoless double beta decay search experiment. RRS can reduce the radon level by a factor of 300, so a high-sensitivity radon detector was required. A highly sensitive radon detector was constructed using a 70 L chamber with a large PIN photodiode to measure radon concentration in the purified air. The radon detector shows an excellent resolution of 72 keV (FWHM) for 6.003 MeV alphas from 218218{}^{218}start_FLOATSUPERSCRIPT 218 end_FLOATSUPERSCRIPTPo decay and a sensitivity down to 23.8 ±plus-or-minus\pm± 2.1 mBq/m33{}^{3}start_FLOATSUPERSCRIPT 3 end_FLOATSUPERSCRIPT with a boil-off N22{}_{2}start_FLOATSUBSCRIPT 2 end_FLOATSUBSCRIPT gas sample. The radon concentration level from the RRS measured by the radon detector was below 0.29Bq/m3{\rm 0.29~{}Bq/m{{}^{3}}}0.29 roman_Bq / roman_m start_FLOATSUPERSCRIPT 3 end_FLOATSUPERSCRIPT with a reduction factor of about 300.

1 Introduction

The underground laboratory, developed to reduce the background caused by cosmic rays, must consider the other background created by the Earth’s surface, radon. Radon is derived from uranium and thorium, commonly present in the Earth’s crust, and can be represented by the radium series from U238superscriptU238{}^{238}Ustart_FLOATSUPERSCRIPT 238 end_FLOATSUPERSCRIPT roman_U and Th232superscriptTh232{}^{232}Thstart_FLOATSUPERSCRIPT 232 end_FLOATSUPERSCRIPT roman_Th decay chains, as shown in Fig. 1.

Refer to caption
Refer to caption
Figure 1: A schematic of Radium series from the U238(top)superscriptU238top{}^{238}U(top)start_FLOATSUPERSCRIPT 238 end_FLOATSUPERSCRIPT roman_U ( roman_top ) and Th232(bottom)superscriptTh232bottom{}^{232}Th(bottom)start_FLOATSUPERSCRIPT 232 end_FLOATSUPERSCRIPT roman_Th ( roman_bottom ) decay chain. For β𝛽\betaitalic_β decay, only the half-life information was included.

Radon is a radioactive noble gas with very low chemical reactivity in general conditions. Because of these characteristics, the concentration of radon cannot be measured directly. However, the radon concentration can be inferred through the daughter nuclide, poloniums. Polonium is a radioactive metal with a positive charge, which allows it to be captured from the air via an electric field. Subsequent stages of radon decay ultimately result in lead production, accompanied by the emission of significant radiation. Radon contamination in the experimental environment presents a substantial background source in underground experiments. Therefore, it is crucial to measure the radon concentration.

Radon emanates from rocks, and as underground experimental facilities are encased within rock formations, there is a potential for radon contamination within the underground experimental facility. In such facilities, radon is removed by radon reduction systems (RRS), which reduce the radon concentration in the air by more than a hundred folds. Commercially available radon detectors cannot measure this low radon concentration. Therefore, a highly sensitive PIN photodiode-based radon detector has been designed and built for use at Yemilab.

Yemilab [1, 2, 3] is the new underground laboratory in Korea that has been operating since the end of 2022, which has 16 research spaces with a total area of 3000m23000superscriptm2\rm 3000\ m^{2}3000 roman_m start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT and a 1000-meter overburden. The schematic of Yemilab is shown in Fig. 2.

Refer to caption
Figure 2: A schematic of the Yemilab experimental areas [1].

Yemilab has a cosmic muon background that is four times lower than the previous underground laboratory, Yangyang Underground Laboratory (Y2L) in Yangyang, Korea [2].

The AMoRE [4, 5] and COSINE [6] experiments, two major physics experiments searching for neutrinoless double β𝛽\betaitalic_β decay and weakly interacting massive particles, are being constructed. Both experiments require extremely low background levels to observe very rare processes. It requires efforts to reduce the detector’s intrinsic background and understand and reduce the background from the environment [7].

Noble gases like radon bind weakly to materials. To effectively adsorb radon in the air, a material with a large surface area is advantageous, and charcoal plays this role. A radon reduction system (RRS) is a device designed to reduce radon in the air by drawing radon-rich air and adsorbing the radon into the charcoal tank. As the charcoal tank’s temperature decreases, the charcoal’s capacity to adsorb the radon increases. Several RRS were developed in several underground laboratories and are still in operation today [8, 9, 10].

The radon detector is designed to measure the concentration of radon that decays to airborne polonium in trapped air in a sampling chamber. To achieve this, a sufficient volume is needed to contain sampled air and form an electric field to collect polonium in that air. From this perspective, a radon detector with a large air sample holding chamber was constructed, similar to radon detectors currently being used in other underground laboratories [11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22].

This paper discusses the radon concentration measurements in the air from the RRS with a highly sensitive radon detector at the Yemilab.

2 Radon Reduction System

ATEKO [23], the most popular RRS manufacturer in the world, has provided the system at many underground facilities [8], including the one at Y2L. However, replacing the most high-pressure part of the system requires an additional certification by the Korea Occupational Safety and Health Agency (KOSHA). To avoid this vital regulation issue in Korea, we purchased an RRS manufactured by a Korean company specializing in ultra-pure water filtration systems.

Figure 3 shows the layout of the RRS for the AMoRE-II at Yemilab. Air from the environment is supplied by the compressor and pressurized in the receiver tank. The absorption dryer can remove the moisture in the air under the 70Csuperscript70C\rm-70^{\circ}C- 70 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPT roman_C of dew point. This moisture-free pressurized air is contained in active charcoal tanks up to 10bars10bars\rm 10\ bars10 roman_bars. The temperature of active charcoal in the tanks has been cooled below 70Csuperscript70C\rm-70^{\circ}C- 70 start_POSTSUPERSCRIPT ∘ end_POSTSUPERSCRIPT roman_C by the air chiller to remove radon from the dried air effectively. A heater and HEPA filter units are installed at the last part of the system to warm up and for dust filtration. The control panel operates all these processes automatically, controlled by a Windows-based program for online operation and monitoring.

Refer to caption
Figure 3: The layout of the custom-made Radon Reduction System. 1: Compressor, 2: Receiver tank, 3: Absorption dryer, 4: Air chiller, 5: Heater and HEPA filter units, 6: Active charcoal tanks, and 7: Control panel.

The RRS aims to reduce the radon concentration below 1/100011000\rm 1/10001 / 1000 that of the air in the Yemilab and supply the radon-free air with a capacity up to 60m3/h60superscriptm3h\rm 60\ m^{3}/h60 roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT / roman_h. A pre-tuning of the system was performed in the factory to optimize the operation conditions for a period longer than six months. It was decided to supply the air at 50m3/h50superscriptm3h\rm 50\ m^{3}/h50 roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT / roman_h to minimize the maintenance effort.

3 Radon Detector

Figure 4 shows the radon detector developed for Yemilab. The detector consists of a 70 L air sampling chamber with a PIN photodiode connected to a ceramic feedthrough mounted on the top flange. The chamber is made of 3 mm thick stainless steel, and the inner surface has been electrically polished to reduce the radon emanating from the chamber wall. The chamber has an inlet and an outlet to measure the air sample, and an additional vacuum pump connection is designed to evacuate the air inside the chamber quickly. The corners of the chamber are curved to form a uniform electric field, a different design from the radon chamber used in the radon monitoring for the KIMS experiment at Y2L [24, 25].

Refer to caption
Refer to caption
Figure 4: The picture (left) and schematic (right) of the radon detector. The measurement volume is the rounded part of the chamber below the flange plate. The measurement volume comprises 3mm3mm\rm 3\ mm3 roman_mm thick stainless steel with an electropolished inner wall. The measurement volume is approximately 70L70L\rm 70\ L70 roman_L. The detector has an air inlet, an air outlet, a port for the vacuum gauge, and a port connected to a vacuum pump to evacuate the internal air. There is a hole in the center of the flange for the feedthrough installation. Above the measurement volume is the front-end electronics board with voltage divider and preamplifier circuits. It is covered with a chamber lid. The chamber lid has three connector sockets, each with SHV, BNC, and power wires for the preamplifier. The outer wall of the chamber is connected to the ground.

A windowless silicon PIN photodiode (S3204-09; Hamamatsu Photonics K.K. [26]) with dimensions of 18mm×18mm18mm18mm18~{}\mathrm{mm}\times 18~{}\mathrm{mm}18 roman_mm × 18 roman_mm is used. A negative voltage is applied to the photodiode to establish an electric potential difference between the photodiode and the grounded inner surface of the chamber. Polonium, a positively charged daughter particle from the radon decay, travels to the photodiode in the chamber. Upon reaching the surface of the photodiode, polonium eventually decays and emits α𝛼\alphaitalic_α particles. The emitted α𝛼\alphaitalic_α particles lose energies in the photodiode’s depletion region to create electron-hole pairs through which a signal is generated.

The air is introduced into the chamber via an oil-free vacuum pump (Rocker 400; Rocker Scientific Co. Ltd. [27]). The dust in the sampled air is removed through a dust filter, which performs the same function as in the commercial radon detector (RAD7; Durridge Company, Inc. [28]). After filtration, the air undergoes dehumidification facilitated by a drying unit, a cylindrical acrylic container with approximately 500 g of desiccant (Drierite desiccant 23005 [29]). The air passing through the desiccant has a relative humidity of 2 to 3%percent\%%RH. These components are seamlessly connected using silicon tubes, as shown in Fig.  5.

Refer to caption
Figure 5: A schematic for the radon measurement setup is shown on the left. The solid (dashed) line is the flow line for the air (electrical connection). Other lines are explained in the legends on the top left. The pictures on the right show each part of the measurement setup, while (h) is the radon detector in Fig. 4.

The signal generated by the photodiode is sent to the front-end electronics board via a feedthrough. Figure 6 shows the circuit diagrams of the front-end electronics. The front-end electronics board has a preamplifier (H4083; Hamamatsu Photonics K.K. [30]) and a high-voltage (HV) divider to supply HV between the photodiode and the ground. There are BNC and SHV sockets for signal and HV, respectively. The front-end electronics board is covered with a Faraday cage to shield the external electromagnetic fields. On the left, a socket connects to the SHV cable, and on the right, there is a socket through which signals are transmitted through the BNC cable. The two holes in the center connect via a feedthrough to the photodiodes in the chamber. On the left side of the BNC connector, a board is designed to mount a preamplifier. This preamplifier is Hamamatsu’s H4083 and is installed under a Faraday cage after soldering to the front-end electronics board to shield the external electromagnetic noise. The power to the preamplifier is supplied by the 3-pin connector at the bottom. A HV supply (N470; CAEN Technologies, Inc.) is used to apply the HV for the radon detector and photodiode, and a low voltage (LV) supply (U8032A; Keysight Technologies, Inc. [31]) to the preamplifier.

Refer to caption
Figure 6: The schematic illustrates the front-end electronics, with the top circuit diagram depicting the HV divider section. This circuit is responsible for grounding the chamber and applying HV and bias to the photodiode. The schematics below detail the preamplifier circuit and its power supply. The photodiode signal goes into the preamplifier’s first terminal for amplification, and the output is from terminal 9. This amplified signal is transmitted to the shaping amplifier through the BNC cable.

Front-end electronics signals are shaped and amplified by a pulse-shaping amplifier (570; ORTEC Inc. [32]). The shaping time used is 2μs2𝜇s\rm 2\ \mu s2 italic_μ roman_s. The shaped signal is finally digitized by the 12-bit flash ADC with a sampling rate of 25 MS/s (NKFADC25; Notice Korea Co.). The pulse height trigger determines whether to save the waveform. Figure 7 shows a typical waveform of an α𝛼\alphaitalic_α signal. All of the analysis parameters are reconstructed from the waveform. Table 1 defines six parameters used in the analysis.

Refer to caption
Figure 7: An example waveform of an α𝛼\alphaitalic_α event after the pedestal subtraction.
Table 1: The definition of parameters reconstructed from waveform.
Parameter Description
PandPRMS𝑃andsubscript𝑃𝑅𝑀𝑆P\ {\rm and}\ P_{RMS}italic_P roman_and italic_P start_POSTSUBSCRIPT italic_R italic_M italic_S end_POSTSUBSCRIPT pedestal mean and RMS
HandHx𝐻andsubscript𝐻𝑥H\ {\rm and}\ H_{x}italic_H roman_and italic_H start_POSTSUBSCRIPT italic_x end_POSTSUBSCRIPT maximum pulse height and its position.
A𝐴Aitalic_A integrated area (HX±2μsplus-or-minussubscript𝐻𝑋2𝜇sH_{X}\pm 2~{}\rm{\mu}sitalic_H start_POSTSUBSCRIPT italic_X end_POSTSUBSCRIPT ± 2 italic_μ roman_s)
Ttrgsubscript𝑇trgT_{\rm{trg}}italic_T start_POSTSUBSCRIPT roman_trg end_POSTSUBSCRIPT trigger time

To correct the baseline of the waveform, the mean of the pedestal is calculated using the data in the 6 μ𝜇\muitalic_μs window before the pulse, and each ADC value of the waveform is subtracted by the mean of the pedestal, as shown in Fig. 7. The parameters corresponding to the energy of an event are H𝐻Hitalic_H and A𝐴Aitalic_A. The H𝐻Hitalic_H is the maximum height of the waveform, and A𝐴Aitalic_A is the total ADC count calculated by integrating the waveform in the range of Hx±2μplus-or-minussubscript𝐻𝑥2𝜇H_{x}\pm 2~{}\muitalic_H start_POSTSUBSCRIPT italic_x end_POSTSUBSCRIPT ± 2 italic_μs. All of these have characteristics proportional to the signal’s size (energy). The parameter H𝐻Hitalic_H is used as the corresponding energy value. The parameter A𝐴Aitalic_A is used primarily for noise rejection. The P𝑃Pitalic_P and PRMSsubscript𝑃RMSP_{\rm{RMS}}italic_P start_POSTSUBSCRIPT roman_RMS end_POSTSUBSCRIPT are calculated at the first 6μs6𝜇s\rm 6\ \mu s6 italic_μ roman_s time window of the waveform. In the case of short-term electrical noise or instability, these parameters tend to deviate from the nominal case. Some electrical noise signals are filtered through these parameters.

As shown in Fig. 8, four peaks are observed. The first three peaks (Po212,218Po,and214Po{}^{212}Po,\ ^{218}Po,\ and\ ^{214}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po , start_POSTSUPERSCRIPT 218 end_POSTSUPERSCRIPT roman_Po , roman_and start_POSTSUPERSCRIPT 214 end_POSTSUPERSCRIPT roman_Po) belong to the U238superscriptU238{}^{238}Ustart_FLOATSUPERSCRIPT 238 end_FLOATSUPERSCRIPT roman_U chain and the last peak is Po212superscriptPo212{}^{212}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po from the Th232superscriptTh232{}^{232}Thstart_FLOATSUPERSCRIPT 232 end_FLOATSUPERSCRIPT roman_Th chain. Each peak, except for Po212superscriptPo212{}^{212}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po, is fitted with the Crystal Ball function [Eq. 3.1], and the fitting results are summarized in Table 2.

f(x;α,n,x¯,σ)={exp((xx¯)22σ2),forxx¯σ>αA(B(xx¯)σ)n,forxx¯σα\centering f(x;\alpha,n,\bar{x},\sigma)=\left\{\begin{matrix}\exp\left(-\frac{% (x-\bar{x})^{2}}{2\sigma^{2}}\right),\ &{\rm for}\ \frac{x-\bar{x}}{\sigma}>-% \alpha\\ A\cdot\left(B-\frac{(x-\bar{x})}{\sigma}\right)^{-n},\ &{\rm for}\ \frac{x-% \bar{x}}{\sigma}\leq-\alpha\end{matrix}\right.\@add@centeringitalic_f ( italic_x ; italic_α , italic_n , over¯ start_ARG italic_x end_ARG , italic_σ ) = { start_ARG start_ROW start_CELL roman_exp ( - divide start_ARG ( italic_x - over¯ start_ARG italic_x end_ARG ) start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT end_ARG start_ARG 2 italic_σ start_POSTSUPERSCRIPT 2 end_POSTSUPERSCRIPT end_ARG ) , end_CELL start_CELL roman_for divide start_ARG italic_x - over¯ start_ARG italic_x end_ARG end_ARG start_ARG italic_σ end_ARG > - italic_α end_CELL end_ROW start_ROW start_CELL italic_A ⋅ ( italic_B - divide start_ARG ( italic_x - over¯ start_ARG italic_x end_ARG ) end_ARG start_ARG italic_σ end_ARG ) start_POSTSUPERSCRIPT - italic_n end_POSTSUPERSCRIPT , end_CELL start_CELL roman_for divide start_ARG italic_x - over¯ start_ARG italic_x end_ARG end_ARG start_ARG italic_σ end_ARG ≤ - italic_α end_CELL end_ROW end_ARG (3.1)

The resolutions of all peaks (σ/x¯𝜎¯𝑥\sigma/\bar{x}italic_σ / over¯ start_ARG italic_x end_ARG) are smaller than 1%. Po218superscriptPo218{}^{218}Postart_FLOATSUPERSCRIPT 218 end_FLOATSUPERSCRIPT roman_Po and Po214superscriptPo214{}^{214}Postart_FLOATSUPERSCRIPT 214 end_FLOATSUPERSCRIPT roman_Po, which are from the same decay chain of Rn222superscriptRn222{}^{222}Rnstart_FLOATSUPERSCRIPT 222 end_FLOATSUPERSCRIPT roman_Rn, are used to estimate the concentration of radon. Due to its long half-life, 138 days, Po210superscriptPo210{}^{210}Postart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Po is unsuitable for estimating the concentration. Pb210superscriptPb210{}^{210}Pbstart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Pb, which follows Po214superscriptPo214{}^{214}Postart_FLOATSUPERSCRIPT 214 end_FLOATSUPERSCRIPT roman_Po, has a half-life of about 22.3years22.3years\rm 22.3\ years22.3 roman_years. Because of this characteristic, Pb210superscriptPb210{}^{210}Pbstart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Pb accumulates in the chamber and generates a signal. The rate of Po210superscriptPo210{}^{210}Postart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Po decay from the Pb210superscriptPb210{}^{210}Pbstart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Pb is proportional to the age of the chamber. To eliminate this, the inside of the chamber must be electropolished, and the silicon diode must be replaced. The Po212superscriptPo212{}^{212}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po of the Th232superscriptTh232{}^{232}Thstart_FLOATSUPERSCRIPT 232 end_FLOATSUPERSCRIPT roman_Th decay chain has a different shape than the other polonium peaks. Due to the short half-life of 300ns300ns\rm 300\ ns300 roman_ns, Po212superscriptPo212{}^{212}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po decays before reaching the photodiode and loses its recoil energy. Therefore, the locations of the three polonium peaks in the U238superscriptU238{}^{238}Ustart_FLOATSUPERSCRIPT 238 end_FLOATSUPERSCRIPT roman_U chain are used for the energy calibration through a linear function, as shown in Fig. 8.

Refer to caption
Figure 8: The ADC distribution with their fitting results. The ADC value is the difference between height and pedestal (HP𝐻𝑃H-Pitalic_H - italic_P) for each event. A total of four peaks are shown in the figure in the order of Po210superscriptPo210{}^{210}Postart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Po, Po218superscriptPo218{}^{218}Postart_FLOATSUPERSCRIPT 218 end_FLOATSUPERSCRIPT roman_Po, Po214superscriptPo214{}^{214}Postart_FLOATSUPERSCRIPT 214 end_FLOATSUPERSCRIPT roman_Po, and Po212superscriptPo212{}^{212}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po. Peaks in the U238superscriptU238{}^{238}Ustart_FLOATSUPERSCRIPT 238 end_FLOATSUPERSCRIPT roman_U chain are fitted with a Crystal Ball function, and the results are listed in Table 2. The upper part of the inset shows the logarithmic scale of the same distribution. As mentioned in the main text, Po212superscriptPo212{}^{212}Postart_FLOATSUPERSCRIPT 212 end_FLOATSUPERSCRIPT roman_Po has a short half-life of 300ns300ns\rm 300\ ns300 roman_ns, which tends to differ from other peaks. The energy calibration plot for each peak is below. The linear function is used.
Table 2: The mean value (x¯¯𝑥\bar{x}over¯ start_ARG italic_x end_ARG) and σ𝜎\sigmaitalic_σ after Crystall Ball function fitting in Fig. 8. The sigma resolution is calculated by σ𝜎\sigmaitalic_σ and x¯¯𝑥\bar{x}over¯ start_ARG italic_x end_ARG, and the FWHM is calculated by the histogram. EαsubscriptE𝛼\rm{E}_{\alpha}roman_E start_POSTSUBSCRIPT italic_α end_POSTSUBSCRIPT, alpha energy, values are from NNDC [33].
Isotopes EαsubscriptE𝛼\rm{E}_{\alpha}roman_E start_POSTSUBSCRIPT italic_α end_POSTSUBSCRIPT [MeV] x¯¯𝑥\bar{x}over¯ start_ARG italic_x end_ARG [ADC] σ𝜎\sigmaitalic_σ [ADC] σ/x¯𝜎¯𝑥\sigma/\bar{x}italic_σ / over¯ start_ARG italic_x end_ARG [%] FWHM [keV]
Po210superscriptPo210{}^{210}Postart_FLOATSUPERSCRIPT 210 end_FLOATSUPERSCRIPT roman_Po 5.304 587.506(33) 4.214(37) 0.72 81
Po218superscriptPo218{}^{218}Postart_FLOATSUPERSCRIPT 218 end_FLOATSUPERSCRIPT roman_Po 6.003 666.205(7) 3.275(5) 0.49 72
Po214superscriptPo214{}^{214}Postart_FLOATSUPERSCRIPT 214 end_FLOATSUPERSCRIPT roman_Po 7.687 857.450(5) 3.376(4) 0.39 72

To calibrate the HV and radon concentration of the radon detector, a closed loop system consisting of a radon detector, RAD7, and radium source is prepared as shown in Fig. 5. The efficiency of the radon detector depends on the HV applied. As the HV increases, the electric field for capturing polonium in the chamber intensifies. The efficiency is maximum when the voltage is high enough to capture all the polonium. RAD7 has a hemispherical detection volume of 0.7L0.7L\rm 0.7\ L0.7 roman_L coated inside with an electrical conductor, and a silicon photodiode is installed at the center of the hemisphere. The radium source, purchased from the Korea Research Institute of Standards and Science (KRISS), has an activity of approximately 1.5kBq1.5kBq\rm 1.5\ kBq1.5 roman_kBq and is sealed within a stainless steel container, as shown in Fig. 5(f). A filter is installed at the container outlet to prevent radium from being blown out directly. Measurements are taken at six points by increasing 250V250V\rm 250\ V250 roman_V each time, from 1000 to 2250V2250V\rm 2250\ V2250 roman_V, and each measurement takes an hour. The activity ratios (R𝑅Ritalic_R) for HV are calculated, and its definition is

R=Ac/AR,𝑅subscript𝐴𝑐subscript𝐴𝑅\centering R=A_{c}/A_{R},\@add@centeringitalic_R = italic_A start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT / italic_A start_POSTSUBSCRIPT italic_R end_POSTSUBSCRIPT , (3.2)

where Acsubscript𝐴𝑐A_{c}italic_A start_POSTSUBSCRIPT italic_c end_POSTSUBSCRIPT is the number of 218218{}^{218}start_FLOATSUPERSCRIPT 218 end_FLOATSUPERSCRIPTPo within 3σ3𝜎3\sigma3 italic_σ range of fitting result on an hour measurement, ARsubscript𝐴𝑅A_{R}italic_A start_POSTSUBSCRIPT italic_R end_POSTSUBSCRIPT is the activity measured by the RAD7. The R𝑅Ritalic_R versus HV is shown in Fig. 9.

Refer to caption
Figure 9: The activity ratios obtained from each data set from 100010001000~{}1000 to 2250V2250V2250~{}\rm V2250 roman_V with a step of 250V250V250~{}\rm V250 roman_V. Each point is measured for about an hour. The location of the red diamond marker is where R𝑅Ritalic_R begins to saturate. This point is set as the operating voltage.

The operating voltage is determined at the point where R𝑅Ritalic_R starts to saturate, 1750V1750V\rm 1750\ V1750 roman_V.

Radon-rich air is injected into the closed-loop configuration to calibrate the detector’s radon concentration. In the closed loop of the radon detector and RAD7, as shown in Fig. 5, the sampled air is circulated with the source for approximately three hours to saturate it sufficiently. The source is isolated from the loop after confirming that the radon concentration is saturated. Radon remaining in the closed-loop structure naturally decays. The radon detector and RAD7 measure the radon concentration in the closed loop simultaneously and continuously, which shows a reduction in radon levels.

Refer to caption
Figure 10: The correlation between RAD7 and radon detector after saturation of radon concentration. Over two weeks, the naturally declining radon concentration is measured using two detectors (RAD7 and the radon detector). A linear calibration function (red) without an offset fits the data.

Figure 10 shows the correlation of radon concentration between RAD7 and the radon detector up to similar-to\sim 300 Bq/m33{}^{3}start_FLOATSUPERSCRIPT 3 end_FLOATSUPERSCRIPT with about two weeks of data. Each point corresponds to data collected for 2 hours. The correlation is fitted by a linear function without an offset, and the calibration function is obtained.

To understand the intrinsic background of the radon detector, about 15L/min15Lmin\rm 15\ L/min15 roman_L / roman_min of pure nitrogen gas from boil-off in a liquid nitrogen gas bottle is flushed to the chamber for 2hours2hours\rm 2\ hours2 roman_hours. The amount of nitrogen gas is enough to refresh the chamber volume approximately 25 times. The humidity within the chamber is about 40ppm40ppm\rm 40\ ppm40 roman_ppm when the chamber is sealed for measurement. Data are collected for approximately 12 days, and the intrinsic background of the radon detector is measured to be 23.8±2.1mBq/m3plus-or-minus23.82.1mBqsuperscriptm3\rm 23.8\pm 2.1\ mBq/m^{3}23.8 ± 2.1 roman_mBq / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT.

4 Radon Concentration Measurements

Environmental monitoring at Yemilab is in progress, including the concentration of radon. Radon-reduced air by the RRS is piped to the AMoRE-II detector room and measured by the radon detector. Ambient air, before RRS, is measured by the RS9A, which is a compact pulsed-ionization type radon detector developed by RadonFTLab in South Korea [34, 35], and can measure the radon concentration below 3.7kBq/m33.7kBqsuperscriptm3\rm 3.7\ kBq/m^{3}3.7 roman_kBq / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT with an accuracy of less than 15%.

Refer to caption
Figure 11: A comparison plot of radon concentration between the experiment hall air and the radon-reduced air from the RRS. Each data is measured by the radon detector (RS9A) for AMoRE’s detector room air from the RRS (AMoRE experimental hall air to the RRS). It is assumed that the uncertainty of the RS9A data set is 15% in precision. The blue dashed line is the AMoRE-II radon concentration requirement.

Figure 11 compares radon concentrations between the AMoRE experimental hall and the radon-reduced air from the RRS. The last nine days of data are used to estimate the reduction factor when the RRS is settled down. The average radon concentration levels before and after RRS are 17.6±0.3plus-or-minus17.60.3\rm 17.6\pm 0.317.6 ± 0.3 and 0.058±0.004Bq/m3plus-or-minus0.0580.004Bqsuperscriptm3\rm 0.058\pm 0.004\ Bq/m^{3}0.058 ± 0.004 roman_Bq / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT, respectively, and the reduction factor is about 1/30013001/3001 / 300. The requirement of radon concentration for AMoRE-II is 0.29Bq/m30.29Bqsuperscriptm3\rm 0.29\ Bq/m^{3}0.29 roman_Bq / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT [36], aligning with the region of interest for the neutrinoless double β𝛽\betaitalic_β decay Q-value of Mo100superscriptMo100{}^{100}Mostart_FLOATSUPERSCRIPT 100 end_FLOATSUPERSCRIPT roman_Mo. Figure 11 shows that the radon concentration satisfies the requirement after ten days.

The calibration process of the radon detector performed in this paper is a relative calibration using a RAD7 in a closed loop. Even though there is a desiccant in the loop, the RAD7 shows the humidity level to be 2 to 3%percent\%%RH, which is about 0.58g/m30.58gsuperscriptm3\rm 0.58\ g/m^{3}0.58 roman_g / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT. Since the humidity of the radon-reduced air from RRS is 0.21g/m3similar-toabsent0.21gsuperscriptm3\sim\rm 0.21\ g/m^{3}∼ 0.21 roman_g / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT, humidity is slightly different between the two setups. However, it seems negligible in calibration constant values between the two when looking at the humidity dependence of calibration constant in Fig. 5 of [13]. The RRS output measurement with lower humidity conditions could be overestimated due to a possible under-correction.

More rigorous calibration works with different humidity environments and flow rates will be carried out in the near future.

5 Summary

The Yemilab, an underground laboratory in Jeongseon, Korea, was built for rare event search experiments to reduce cosmic-ray backgrounds. However, the Yemilab is still impacted by the radon background in the air.

An RRS using a cold and pressurized charcoal chamber was designed and installed at the Yemilab to reduce the background radon in the AMoRE experiment, which is looking for the neutrinoless double beta decay from 100100{}^{100}start_FLOATSUPERSCRIPT 100 end_FLOATSUPERSCRIPTMo isotopes in lithium molybdate crystal detectors at cryogenic temperatures. It reduces the radon level in the air at the experimental area to the level where the radon background becomes negligible to the experiment’s sensitivity with a capacity of 50m3/h50superscriptm3h\rm 50\ m^{3}/h50 roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT / roman_h.

A highly sensitive radon detector with a 70 L chamber volume and a large silicon PIN photo-diode sensor readout was designed and manufactured to quantify the amount of radon reduced by the RRS. A calibration function for different HV values up to 2250 V was obtained using a closed loop connected with a calibrated commercial radon detector RAD7 using a radium source. Also, a radon activity calibration using the same setup was made up to similar-to\sim 300 Bq/m33{}^{3}start_FLOATSUPERSCRIPT 3 end_FLOATSUPERSCRIPT at 1750 V. From the calibration, the radon detector can measure alpha particles from Po218superscriptPo218{}^{218}Postart_FLOATSUPERSCRIPT 218 end_FLOATSUPERSCRIPT roman_Po (6.003 MeV) and Po214superscriptPo214{}^{214}Postart_FLOATSUPERSCRIPT 214 end_FLOATSUPERSCRIPT roman_Po (7.687 MeV), daughter isotopes of Rn222superscriptRn222{}^{222}Rnstart_FLOATSUPERSCRIPT 222 end_FLOATSUPERSCRIPT roman_Rn, with an FWHM resolution of 72 keV. A boil-off N22{}_{2}start_FLOATSUBSCRIPT 2 end_FLOATSUBSCRIPT gas test showed 23.8±2.1mBq/m3plus-or-minus23.82.1mBqsuperscriptm3\rm 23.8\pm 2.1\ mBq/m^{3}23.8 ± 2.1 roman_mBq / roman_m start_POSTSUPERSCRIPT 3 end_POSTSUPERSCRIPT, the lowest background level indicating an intrinsic background.

To confirm the reduction in radon by RRS, air from the RRS was measured using the radon detector, while the air going into the RRS was measured by another commercial radon detector, RS9A, for 19 days. From the measurement, a radon reduction factor of similar-to\sim 300 by the RRS was measured. Also, the RRS output level reached below 0.29 Bq/m33{}^{3}start_FLOATSUPERSCRIPT 3 end_FLOATSUPERSCRIPT, which is the radon concentration requirement of AMoRE-II.

Acknowledgments

This work was supported by the Institute for Basic Science (IBS) funded by the Ministry of Science and ICT, Korea (Grant: IBS-R016-D1).

References

  • [1] K. S. Park et al., The new underground facility in Korea, Yemilab, Journal of Physics: Conference Series 2156 012171 (2021).
  • [2] K. S. Park et al., Construction of Yemilab, Front. Phys. 12 1323991 (2024).
  • [3] M. H. Lee Radioassay and Purification for Experiments at Y2L and Yemilab in Korea, Journal of Physics: Conference Series 1468 012249 (2020).
  • [4] V. Alenkov et al. (AMoRE collaboration), Technical Design Report for the AMoRE 0νββ0𝜈𝛽𝛽0\nu\beta\beta0 italic_ν italic_β italic_β Decay Search Experiment, arxiv:1512.05957 (2015).
  • [5] M. H. Lee (AMoRE collaboration), AMoRE: a search for neutrinoless double-beta decay of 100100{}^{\rm 100}start_FLOATSUPERSCRIPT 100 end_FLOATSUPERSCRIPTMo using low-temperature molybdenum-containing crystal detectors, JINST 15(08) C08010 (2020).
  • [6] G. Adhikari et al. (COSINE collaboration), Initial performance of the COSINE-100 experiment, The European Physical Journal C 78 107 (2018).
  • [7] C. Ha et al., Radon concentration variations at the Yangyang underground laboratory, Front. Phys. 10 1030024 (2022).
  • [8] J. Pérez-Pérez et al., Radon Mitigation Applications at the Laboratorio Subterráneo de Canfranc (LSC), Universe 8(2) 112 (2022).
  • [9] R. Hodák et al., Characterization and long-term performance of the Radon Trapping Facility operating at the Modane Underground Laboratory, Journal of Physics G: Nuclear and Particle Physics 46(11) 115105 (2019).
  • [10] M. Murra et al., Design, construction and commissioning of a high-flow radon removal system for XENONnT, The European Physical Journal C 82(12) 1104 (2022).
  • [11] Y. Takeuchi et al., Development of high sensitivity radon detectors, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 421 334 (1999).
  • [12] J.-X. Wang et al., An electrostatic radon detector designed for water radioactivity measurements, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 421 601 (1999).
  • [13] K. Hosokawa et al., Development of a high-sensitivity 80 L radon detector for purified gases, Progress of Theoretical and Experimental Physics 2015 033H01 (2015).
  • [14] C. Mitsuda et al., Development of super-high sensitivity radon detector for the Super-Kamiokande detector, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 497 414 (2003).
  • [15] S. Folea et al., WI-FI portable solution for distributed radon measurements, Romanian Journal of Physics 58 S126 (2013).
  • [16] J.L. Gutiérrez et al., Radon emanation measurements using silicon photodiode detectors, Applied Radiation and Isotopes 60 583 (2004).
  • [17] K. Okamoto et al., Improvement of radon detector performance by using a large-sized PIN-photodiode, arxiv:2112.06614 (2021).
  • [18] S. Elisio and L. Peralta, Development of a low-cost monitor for radon detection in air, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 969 164033 (2020).
  • [19] E.A. Yakushev and A.V. Lubashevskii, Radon-induced background and methods of its elimination in the EDELWEISS-II experiment, Russian Physics Journal 53 616 (2010).
  • [20] E. Choi et al., Highly sensitive radon monitor and radon emanation rates for detector components, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 459 177 (2001).
  • [21] J. Kiko, Detector for R222nsuperscript𝑅222𝑛{}^{222}Rnstart_FLOATSUPERSCRIPT 222 end_FLOATSUPERSCRIPT italic_R italic_n measurements in air at the 1 mBq/m33{}^{3}start_FLOATSUPERSCRIPT 3 end_FLOATSUPERSCRIPT level, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 460 272 (2001).
  • [22] F. Mamedov et al., Development of an ultra-sensitive radon detector for the SuperNEMO experiment, JINST 6(12) C12008 (2011).
  • [23] ATEKO a.s., https://www.ateko.cz/en, (Accessed February 15, 2024).
  • [24] M. Lee et al., Radon Environment in the Korea Invisible Mass Search Experiment and Its Measurement, Journal of the Korean Physical Society 58 713 (2011).
  • [25] K. M. Seo et al., Silicon PIN photodiode-based radon detectors for an underground experiment environment, Proceedings of The 39th International Conference on High Energy Physics — PoS(ICHEP2018) 340 803 (2019).
  • [26] Hamamatsu silicon photodiode S3204-09, https://www.hamamatsu.com/us/en/product/optical-sensors/photodiodes/si-photodiodes/S3204-09.html, (Accessed February 16, 2024).
  • [27] Rocker 400 vacuum pump, https://www.rocker.com.tw/en/product/lab-pumps/oil-free-vacuum-pump/rocker-400-oil-free-vacuum-pump/, (Accessed February 16, 2024).
  • [28] RAD7 RADON DETECTOR, https://durridge.com/products/rad7-radon-detector/, (Accessed February 16, 2024).
  • [29] DRIERITE 23005, https://secure.drierite.com/catalog3/product.cfm?item_num=23005/, (Accessed February 16, 2024).
  • [30] Hamamatsu charge amplifier H4083, https://www.hamamatsu.com/content/dam/hamamatsu-photonics/sites/documents/99_SALES_LIBRARY/ssd/charge_amp_kacc9001e.pdf, (Accessed February 16, 2024).
  • [31] Keysight Triple Output DC Power Supply U8032A, https://www.keysight.com/us/en/product/U8032A/triple-output-dc-power-supply-60v-3a-2x-5v-3a-375w.html, (Accessed February 16, 2024).
  • [32] Ortec amplifier 570, https://www.ortec-online.com/-/media/ametekortec/brochures/5/570.pdf?la=en&revision=53f9a77c-d8f8-4459-80b8-a93822b2aab2, (Accessed February 16, 2024).
  • [33] National Nuclear Data Center, https://www.nndc.bnl.gov/, (Accessed February 15, 2024).